Recensioner

0 recensioner

Du kanske även behöver

Milwaukee MW600 PRO Syremätare med ATC

Milwaukee MW600 PRO Syremätare med ATC

Milwaukee MW600 PRO Syremätare är en snabb och pålitlig mätare för mätning av löst syre (Dissolved Oxygen) i vätskor. Förkortas ofta DO-meter eller DO-mätare. Med automatisk temperaturkorrigering (ATC) mellan 0-30°C.Milwaukee MW600 PRO Syremätare med ATC- Polarografisk sond med PTFE-sensor och 3 meter kabel- Automatisk temperaturkorrigering 0-30°C- Kalibreringsskruvmejsel- Membran och o-ringar, 5 st- Elektrolytlösning, 30 ml- 9V batteri- Instruktion på engelskaOBS! Du behöver "Milwaukee MA9070 Solution" för att kunna kalibrera mätaren!---Milwaukee MW600 PRO Dissolved Oxygen Meter med ATC- Fast, easy-to-read results with large digital LCD readout- Repeatable, accuracy to ±1.5% FS- Lightweight and portable- Simple 2 point open air calibration (see video below)- Uses MA840 reinforced polarographic probe with an oxygen-permeable polytetrafluroethylene (PTFE) membrane sensor- 5 o-rings, membranes and one protective cap included- 10 feet probe cable for extra flexibility when testing- 70 hours of battery life with continuous use (battery included) with low battery warningSPECIFICATIONSO2 Range: 0.0 to 19.9 mg/LO2 resolution: 0.1 mg/LO2 Accuracy: ±1.5% Full ScaleCalibration: Manual on 2 points (zero and slope)Temperature Compensation: Automatic, from 0 to 30°CProbe: MA840 refillable polarographic probe includedMembranes: 5 includedEnvironment: 0 to 50°C / 32 to 122°F; max RH 95%Battery Type: 1 x 9V alkaline (included)Battery Life: Approx. 70 hours of continuous useDimensions: 5.7 x 3.1 x 1.6 inchesWeight: 7.5 ounces with batteryCARE AND USE“OPEN AIR” CALIBRATION PROCEDURE1. Remove the membraneTake the membrane off the end of the probe and dry the diode off.2. Zero out the displayConnect the probe to the base unit and turn the unit on. Your display should read 0.0. If it does not, go to the position “0” screw on the front of the unit with the screw driver provided and through the access hole make the unit display read 0.0.3. Fill the membraneGo to the membrane and put the “O” ring inside the membrane on the shoulder at the base of the threads. Fill the membrane to just short of the “O” ring with MA9071 electrolyte solution. Be sure that there are no bubbles inside the membrane. (Air bubbles are the main cause of calibration failure)4. Attach the membraneScrew the membrane on to the probe.5. Let stablizeTurn the meter on and hang the probe off the end of a table pointing down and let the unit display stabilize (this may take 5 to 15 minutes)6. AdjustPush down and hold the “Cal” button on the front face of the meter and with the screwdriver go to the front Cal position “100%” and make the display read 100.You are now calibrated.EXTRA TIPS- Watch for the membrane having wrinkles or stretch marks. If you see wrinkles, replace with a new membrane.- Always use fresh electrolyte solution (MA9071).

2 199 kr